Human Magnetic Luminex® Performance Assay MMP Base Kit

Build your own Luminex Assay with our Luminex Assay Customization Tool.

U.S. orders usually ship in 2-3 business days

Build Now
Human Magnetic Luminex Performance Assay MMP Base Kit
1 Image
Product Details
Citations (12)
FAQs
Supplemental Products
Reviews (2)

Human Magnetic Luminex® Performance Assay MMP Base Kit Summary

Panel Luminex Performance Human MMP Panel
Assay Type Magnetic bead-based multiplex assay for the Luminex® platform
Format 1 x 96-well microplate and Magnetic antibody-coated beads
Species Human
Analytes Detected Please see analyte list in assay customization tool below.
Performance Validation Our Luminex High Performance Assays undergo our most extensive validation testing. All Luminex High Performance Assay Panels are validated for use with serum and plasma samples. Some High Performance Assays are additionally validated for cell culture supernatants, milk, saliva, or urine, as indicated on the individual product data sheets. Luminex High Performance Assays are tested for sensitivity (three-quarters the low standard), intra-assay precision, inter-assay precision, and to ensure assay linearity for validated sample types. Recovery values for individual samples in validated sample types are also tested. Validation data for each analyte can be found in the product datasheet.

Make your selections below to configure your assay

Filter Analytes
Select All Select None

Assays

Below are kits that are available based on your selections. These selections may be further customized before adding to the cart or emailing the configuration to your purchasing department or local distributor to place your order. Selected analytes may be included in more than one High Performance Assay. All assay options are displayed.

If you run into bead region incompatibilities, send your Luminex code to our team and find out more.

How would you like these analytes mixed?

Premixed
User Mixed
Please select at least one mixing option to see the kits.

Would you like to split based on dilution?

Do not split
Split based on serum
Split based on plasma
Split based on cell culture supernates

Not finding your analyte of interest? View our complete Luminex offering in the full version of our Luminex Assay Customization Tool. We also offer Custom Luminex Services.

Product Features

Magnetic Luminex High Performance Assays are flexible bead-based multiplex assays. They allow up to 50 user-defined target analytes to be simultaneously profiled using cell culture supernates, serum, or plasma samples. Some panels are also validated with human milk, saliva, and/or urine samples.

  • Simultaneously profile up to analytes of your choice in one sample
  • Magnetic format allows for easier wash steps
  • Undergoes similar validation testing as Quantikine® ELISA Kits
  • Precise and Specific
  • Requires a small sample volume (<50 µL)
  • Run your assay in just 3-5 hours
  • Mass-calibrated standards for consistent results with every new lot of material
  • Enough reagents for one 96-well plate
Base Kit Components
  • Standard Cocktail(s)
  • Bead Diluent
  • Biotin Antibody Diluent
  • Calibrator Diluent
  • Wash Buffer
  • Streptavidin-PE
  • One flat-bottom 96-well Microplate
  • Foil Plate Sealers (4)
  • Mixing Bottle
  • Standard Value Card**

**A standard curve must be generated each time an assay is run, utilizing values from the Standard Value Card included in the Base Kit.

Bead Set Contents
  • User-mix antibody-coated Magnetic beads in individual vials
  • User-mixed cocktail of biotinylated detection antibody

Precision

Intra-Assay Precision -  is generated from the mean of the %CV's from 60 reportable results across three different concentrations of analytes in a single serum assay.

Inter-Assay Precision - is generated from the mean of the %CV's across three different concentrations of analytes across 72 different serum assays.

  n=60 (20x3 levels) n=72
Analyte Intra-Assay (%CV) Inter-Assay (%CV)
EMMPRIN/CD147 6.5 9.8
MMP-1 8.7 15.8
MMP-2 7.9 11.6
MMP-3 6 10.3
MMP-7 7 9.8
MMP-8 6.1 11.2
MMP-9 5 10.5
MMP-10 6.1 10.8
MMP-12 7.9 10.8
MMP-13 5 11.8

Accuracy

Linearity - The data represents mean spiked or natural linearity in serum matrix samples.

Spike Recovery - The data represents mean percent recovery of spiked standards ranging from low, medium, and high concentration in serum matrix samples.

  Linearity Spiked
Recovery
Analyte Average % Average %
EMMPRIN/CD147 81 105
MMP-1 109 NO
MMP-2 103 106
MMP-3 99 NO
MMP-7 96 104
MMP-8 110 85
MMP-9 115 NO
MMP-10 100 105
MMP-12 109 87
MMP-13 107 111
Which Luminex Assay is right for me?

Assays for the Luminex platform are offered as High Performance Assays or Assays. The Luminex High Performance Assays are fully validated panels with a focused selection of analytes. They are available in We Mix, You Mix, and Predetermined formats. The Luminex Assays allows for the maximum number of analytes in a multiplex and is supplied as a premixed kit. View a table comparing the features and benefits of these bead-based multiplex assays.

More Information About Magnetic Bead-based Multiplex Kits For The Luminex® Platform

Magnetic Luminex High Performance Assays are designed for use with the Luminex MAGPIX CCD Imager. Alternatively, kits can be used with the Luminex 100/200™ or FLEXMAP 3D, dual laser, flow-based sorting and detection platforms.

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards, and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody cocktail specific to the analytes of interest is added to each well. Following a wash to remove any unbound biotinylated antibody, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE), which binds to the biotinylated detection antibodies, is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer. One laser is bead-specific and determines which analyte is being detected. A magnet in the analyzer captures and holds the superparaMagnetic microparticles in a monolayer. Two spectrally distinct Light Emitting Diodes (LEDs) illuminate the beads. One LED identifies the analyte that is being detected and the second LED determines the magnitude of the PE-derived signal, which is in direct proportion to the amount of analyte bound. Each well is imaged with a CCD camera. Kits can also be used with Luminex 100/200 or a Bio-Rad Bio-Plex dual laser, flow-based systems.

Product Datasheets

You must select a language.

x
Product Specific Notices
Luminex® is a registered trademark of Luminex Corporation.

Citations for Human Magnetic Luminex® Performance Assay MMP Base Kit

R&D Systems personnel manually curate a database that contains references using R&D Systems products. The data collected includes not only links to publications in PubMed, but also provides information about sample types, species, and experimental conditions.

12 Citations: Showing 1 - 10
Filter your results:

Filter by:

  1. Human apical-out nasal organoids reveal an essential role of matrix metalloproteinases in airway epithelial differentiation
    Authors: Li, L;Jiao, L;Feng, D;Yuan, Y;Yang, X;Li, J;Jiang, D;Chen, H;Meng, Q;Chen, R;Fang, B;Zou, X;Luo, Z;Ye, X;Hong, Y;Liu, C;Li, C;
    Nature communications
    Species: Human
    Sample Types: Organoid Culture
  2. Increased Matrix Metalloproteinase-2 and Matrix Metalloproteinase-3 Concentrations in Corneal Epithelium of Patients with Recurrent Corneal Erosions
    Authors: K Jadczyk-So, W Garczorz, B Buba?a-Sta, T Francuz, E Mrukwa-Kom
    Journal of Ophthalmology, 2022-09-25;2022(0):5024037.
    Species: Human
    Sample Types: Cell Lysates
  3. Effect of BCG vaccination on proinflammatory responses in elderly individuals
    Authors: N Pavan Kuma, C Padmapriya, A Rajamanick, SB Marinaik, A Nancy, S Padmanaban, N Akbar, M Murhekar, S Babu
    Science Advances, 2021-08-04;7(32):.
    Species: Human
    Sample Types: Plasma
  4. Evaluation of serum MMP-2 and MMP-3, synovial fluid IL-8, MCP-1, and KC concentrations as biomarkers of stifle osteoarthritis associated with naturally occurring cranial cruciate ligament rupture in dogs
    Authors: S Malek, HY Weng, SA Martinson, MC Rochat, R Béraud, CB Riley
    PLoS ONE, 2020-11-19;15(11):e0242614.
    Species: Canine
    Sample Types: Serum
  5. Characterization of the local wound environment following treatment of chronic leg ulcers with SIS wound matrix
    Authors: JP Hodde, MC Hiles, DW Metzger
    J Tissue Viability, 2019-12-13;0(0):.
    Species: Human
    Sample Types: Wound Fluid
  6. Plasma levels of C-reactive protein, matrix metalloproteinase-7 and lipopolysaccharide-binding protein distinguish active pulmonary or extrapulmonary tuberculosis from uninfected controls in children
    Authors: VVS Albuquerqu, NP Kumar, KF Fukutani, B Vasconcelo, MB Arriaga, PS Silveira-M, S Babu, BB Andrade
    Cytokine, 2019-07-09;123(0):154773.
    Species: Human
    Sample Types: Plasma
  7. Effects of aqueous suppressants and prostaglandin analogues on early wound healing after glaucoma implant surgery
    Authors: KI Jung, JE Woo, CK Park
    Sci Rep, 2019-03-27;9(1):5251.
    Species: Rabbit
    Sample Types: Aqueous Humor
  8. Marked infiltration of neutrophils at the upper palpebral conjunctiva in patients with chronic graft-versus-host disease
    Authors: YS Byun, YS Yoo, MJ Kang, WJ Whang, KS Na, JW Mok, CK Joo
    Ocul Surf, 2018-12-20;0(0):.
  9. Glucocorticoid sensitivity and inflammatory status of peripheral blood mononuclear cells in patients with coronary artery disease
    Authors: S Jönsson, AK Lundberg, RWS Chung, L Jonasson
    Ann. Med., 2018-02-28;0(0):1-24.
    Species: Human
    Sample Types: Plasma
  10. The impact of human adipose tissue-derived stem cells on breast cancer cells: implications for cell-assisted lipotransfers in breast reconstruction
    Authors: E Koellenspe, LC Bonnert, I Zoernig, F Marmé, S Sandmann, G Germann, F Gramley, U Leimer
    Stem Cell Res Ther, 2017-05-25;8(1):121.
    Species: Human
    Sample Types: Cell Culture Supernates
  11. Tumor Necrosis Factor-alpha, Matrix-Metalloproteinases 8 and 9 Levels in the Saliva Are Associated with Increased Hemoglobin A1c in Type 1 Diabetes Subjects.
    Authors: Kuehl M, Rodriguez H, Burkhardt B, Alman A
    PLoS ONE, 2015-04-27;10(4):e0125320.
    Species: Human
    Sample Types: Saliva
  12. Overexpression of MMP-9 and its inhibitors in blood mononuclear cells after myocardial infarction--is it associated with depressive symptomatology?
    Authors: Jonsson S, Lundberg A, Jonasson L
    PLoS ONE, 2014-08-25;9(8):e105572.
    Species: Human
    Sample Types: Cell Culture Supernates

FAQs

No product specific FAQs exist for this product, however you may

View all Luminex Assay FAQs
Loading...

Reviews for Human Magnetic Luminex® Performance Assay MMP Base Kit

Average Rating: 4.5 (Based on 2 Reviews)

5 Star
50%
4 Star
50%
3 Star
0%
2 Star
0%
1 Star
0%

Have you used Human Magnetic Luminex® Performance Assay MMP Base Kit?

Submit a review and receive an Amazon gift card.

$25/€18/£15/$25CAN/¥75 Yuan/¥2500 Yen for a review with an image

$10/€7/£6/$10 CAD/¥70 Yuan/¥1110 Yen for a review without an image

Submit a Review

Filter by:


Human Magnetic Luminex Performance Assay MMP Base Kit
By Anonymous on 10/04/2020
Sample Tested: Serum

Human Magnetic Luminex Performance Assay MMP Base Kit
By Anonymous on 10/19/2019
Sample Tested: in vitro cell culture